GF-M3000 Elisa Analyzer
I. Features:
◆ Big screen LCD display, English graphical interface, touch screen input ,choosing to equip outlay mouse and keyboard, simple and easy for operation
◆ The whole-new operation system ,with integrated software of immunity analysis, can make direct Chinese input of hospital management and patients’ information
◆ 8-channel optical fiber high speed scanning ,special center orientation for elisa hole leads to quick and correct measuring result
◆ Single-board multi-item testing , even 12 kinds qualitative or measurable testing can be done simultaneously on the same board,.
◆ Complete quality-control way ,owning the function of LJ method, X,S method and instant method and double quality-control function by multi-rule, can display, store, print quality-control data and quality-control chart of each item
◆ Visible and flexible board collocation, freely setting of position and quantity of blank , comparison, standard and sample
◆ The instruments can intelligently judge the state of blank, comparison and standard according to the requirement of the program parameter
◆ String database can store 500 program parameters,200 thousand testing results, and more than 2 years of quality-control data and quality-control chart, and freely display and print as well
◆ Automatic single/double wavelength examination ,support many kinds of qualitative or measurable analytical models。
◆ Support many kinds of built-in and outlay printers, print the whole report and quality-control and patient general report
◆ With the function of all-scale query and statistics, convenient for management of sections
◆ The intelligent resting protection function for prolonging the life-span of lamp
◆ The Elisa workstation software can be chosen to equip with as well
II. Main technical index:
◆ Form of sample: 96 or 48 hole Elisa board of strip
◆ Model of calculation: absorbency, cut-off qualitative calculation, single-point scaling , polyline regression ,more-point percentage, linear regression ,index regression and log-logit regression
◆ Light filter: standard configuration with 4 light filters of 405、450、492、630nm,can reach to 8light filters(or according to client’ requirement)
◆ Absorbance accurancy:±0.01A
◆ Absorbance repetition:≤0.5%
◆ Absorbance stability: ±0.005A
◆ Testing speed : 5 seconds/96 holes
◆ Linear error :±0.5%
◆ Display: LCD display
◆ Way of input :touching screen and pen
◆ Interface: serial port RS232, parallel port with printer